Title: | Measurement and analysis of electromagnetic fields of pulsed magnetic field therapy systems for private use |
Authors: | Järmann, Thomas Suter, Fabian Osterwalder, Diego Lüchinger, Roger |
DOI: | 10.1088/0952-4746/31/1/007 |
Published in: | Journal of Radiological Protection |
Volume(Issue): | 31 |
Issue: | 1 |
Page(s): | 107 |
Pages to: | 116 |
Issue Date: | 2011 |
Publisher / Ed. Institution: | IOP Publishing |
ISSN: | 0952-4746 1361-6498 |
Language: | English |
Subject (DDC): | 530: Physics 615: Pharmacology and therapeutics |
Abstract: | Recently, pulsed magnetic field therapy (PMFT) systems have become available for private use. Although they may be applied without medical supervision, only a little is known about their field quantities. In this study, the spatial distribution and the temporal characteristics of the magnetic flux densities of three PMFT systems, available in Europe, were analysed. In close proximity to the surface, the maxima of the peak magnetic flux densities were 461 µT, 170 µT and 133 µT, respectively. At a distance of 30 cm above the whole body mat, the peak magnetic flux density was 77 µT. The excitation patterns consisted of repeating bursts with carrier frequencies between 210 and 1667 Hz. In conclusion, magnetic flux densities were far above International Commission on Non-Ionizing Radiation Protection reference levels. Since these systems are supposed to be medical devices as well as wellness devices, risk analysis of PMFT systems and the effectiveness of these devices need to be investigated in future studies. |
URI: | https://digitalcollection.zhaw.ch/handle/11475/4989 |
Fulltext version: | Published version |
License (according to publishing contract): | Licence according to publishing contract |
Departement: | School of Engineering |
Appears in collections: | Publikationen School of Engineering |
Files in This Item:
There are no files associated with this item.
Show full item record
Järmann, T., Suter, F., Osterwalder, D., & Lüchinger, R. (2011). Measurement and analysis of electromagnetic fields of pulsed magnetic field therapy systems for private use. Journal of Radiological Protection, 31(1), 107–116. https://doi.org/10.1088/0952-4746/31/1/007
Järmann, T. et al. (2011) ‘Measurement and analysis of electromagnetic fields of pulsed magnetic field therapy systems for private use’, Journal of Radiological Protection, 31(1), pp. 107–116. Available at: https://doi.org/10.1088/0952-4746/31/1/007.
T. Järmann, F. Suter, D. Osterwalder, and R. Lüchinger, “Measurement and analysis of electromagnetic fields of pulsed magnetic field therapy systems for private use,” Journal of Radiological Protection, vol. 31, no. 1, pp. 107–116, 2011, doi: 10.1088/0952-4746/31/1/007.
JÄRMANN, Thomas, Fabian SUTER, Diego OSTERWALDER und Roger LÜCHINGER, 2011. Measurement and analysis of electromagnetic fields of pulsed magnetic field therapy systems for private use. Journal of Radiological Protection. 2011. Bd. 31, Nr. 1, S. 107–116. DOI 10.1088/0952-4746/31/1/007
Järmann, Thomas, Fabian Suter, Diego Osterwalder, and Roger Lüchinger. 2011. “Measurement and Analysis of Electromagnetic Fields of Pulsed Magnetic Field Therapy Systems for Private Use.” Journal of Radiological Protection 31 (1): 107–16. https://doi.org/10.1088/0952-4746/31/1/007.
Järmann, Thomas, et al. “Measurement and Analysis of Electromagnetic Fields of Pulsed Magnetic Field Therapy Systems for Private Use.” Journal of Radiological Protection, vol. 31, no. 1, 2011, pp. 107–16, https://doi.org/10.1088/0952-4746/31/1/007.
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.