Publikationstyp: Beitrag in wissenschaftlicher Zeitschrift
Art der Begutachtung: Peer review (Publikation)
Titel: A helium beam path for an imaging-plate detector system
Autor/-in: Estermann, Michael A.
Scheidegger, Stephan
Reifler, Hans
Steurer, Walter
DOI: 10.1107/S0021889897005773
Erschienen in: Journal of Applied Crystallography
Band(Heft): 1997
Heft: 30
Seite(n): 1165
Seiten bis: 1166
Erscheinungsdatum: 1997
Verlag / Hrsg. Institution: Wiley
ISSN: 0021-8898
Sprache: Englisch
Schlagwörter: Imaging plate; Helium beam path; Diffraction
Fachgebiet (DDC): 540: Chemie
Zusammenfassung: A helium beam path for the marresearch imaging plate detector system (X-ray Research GmbH, Norderstedt, Germany) is described. The helium beam path is designed to minimize unwanted air scattering of the primary beam and to reduce absorption of the diffracted beam. The standard marresearch detector system, consisting of the imaging plate detector (online scanner) and the detector base with integrated collimation system, remains unchanged. The beam path consists of a polyethylene bag with the open end fixed to a metal ring. The ring is pressed against the entry window of the imaging plate detector. In this way a sealed volume is created. The primary beam enters into the helium bag through an additional collimating system which is sealed with a thin polyimide window. The crystal and the goniometer head are located inside the helium bag and are connected to the outside spindle axis via a linear and rotary motion feedthrough.
URI: https://digitalcollection.zhaw.ch/handle/11475/3523
Volltext Version: Publizierte Version
Lizenz (gemäss Verlagsvertrag): Lizenz gemäss Verlagsvertrag
Departement: School of Engineering
Enthalten in den Sammlungen:Publikationen School of Engineering

Dateien zu dieser Ressource:
Es gibt keine Dateien zu dieser Ressource.
Zur Langanzeige
Estermann, M. A., Scheidegger, S., Reifler, H., & Steurer, W. (1997). A helium beam path for an imaging-plate detector system. Journal of Applied Crystallography, 1997(30), 1165–1166. https://doi.org/10.1107/S0021889897005773
Estermann, M.A. et al. (1997) ‘A helium beam path for an imaging-plate detector system’, Journal of Applied Crystallography, 1997(30), pp. 1165–1166. Available at: https://doi.org/10.1107/S0021889897005773.
M. A. Estermann, S. Scheidegger, H. Reifler, and W. Steurer, “A helium beam path for an imaging-plate detector system,” Journal of Applied Crystallography, vol. 1997, no. 30, pp. 1165–1166, 1997, doi: 10.1107/S0021889897005773.
ESTERMANN, Michael A., Stephan SCHEIDEGGER, Hans REIFLER und Walter STEURER, 1997. A helium beam path for an imaging-plate detector system. Journal of Applied Crystallography. 1997. Bd. 1997, Nr. 30, S. 1165–1166. DOI 10.1107/S0021889897005773
Estermann, Michael A., Stephan Scheidegger, Hans Reifler, and Walter Steurer. 1997. “A Helium Beam Path for an Imaging-Plate Detector System.” Journal of Applied Crystallography 1997 (30): 1165–66. https://doi.org/10.1107/S0021889897005773.
Estermann, Michael A., et al. “A Helium Beam Path for an Imaging-Plate Detector System.” Journal of Applied Crystallography, vol. 1997, no. 30, 1997, pp. 1165–66, https://doi.org/10.1107/S0021889897005773.


Alle Ressourcen in diesem Repository sind urheberrechtlich geschützt, soweit nicht anderweitig angezeigt.