Please use this identifier to cite or link to this item: https://doi.org/10.21256/zhaw-29026
Full metadata record
DC FieldValueLanguage
dc.contributor.authorGrossmann Aybar, Christian-
dc.contributor.authorFehr, Daniel-
dc.contributor.authorSpano, Fabrizio-
dc.contributor.authorKourouklis, Andreas P.-
dc.contributor.authorMazza, Edoardo-
dc.contributor.authorBonmarin, Mathias-
dc.date.accessioned2023-11-09T11:34:42Z-
dc.date.available2023-11-09T11:34:42Z-
dc.date.issued2023-09-13-
dc.identifier.urihttps://digitalcollection.zhaw.ch/handle/11475/29026-
dc.description.abstractDifferent engineered approaches have led the design of implants with controlled physical features to minimize adverse effects in biological tissues. Aiming to prevent infection, similar efforts have focused on optimizing the design features of drivelines used to transfer power to percutaneous ventricular assistance devices (VAD), omitting however a thorough look on the implant-skin interactions that govern local tissue reactions. Here, we utilized an integrated approach for the biophysical modification of transdermal implants and their evaluation by chronic sheep implantation in comparison to the standard of care VAD drivelines. We developed a novel method for the transfer of breath topographical features on thin wires with modular size. Moreover, we examined the impact of implant’s diameter, surface topography, and chemistry on macroscopic, histological, and physical markers of inflammation, fibrosis, and mechanical adhesion. All implants demonstrated infection-free performance. The fibrotic response was enhanced by the increasing diameter of implants but not influenced by their surface properties. The implants of 0.2 mm diameter promoted mild inflammatory responses with improved mechanical adhesion and restricted epidermal downgrowth, in both silicone and polyurethane coated transdermal wires. On the contrary, the VAD drivelines with larger diameter triggered severe inflammatory reactions with frequent epidermal downgrowth [1]. Furthermore, we performed COMSOL simulations to investigate the electrothermal implications of conductive wires with different sizes for the power transfer to VADs. In our model, we simulate the electrical properties of the prototype’s wires, to confirm that it does not produce a significant body temperature rise. The skin model (Fig. 1a) mimics the multiple skin layer’ properties of epidermis, dermis, fat and muscle [2]. Also, we include a PDMS layer (5 mm thickness) that represents the silicon-based material of the conductive skin. During the study, different thicknesses of the polyurethane (PU) insulating coating were tested for the wire of 0.2 mm diameter. However, no significant improvement was observed when increasing the insulation layer, since the temperature difference in the model was due to the temperature skin gradient and not the electric current. In this model, the dimensions of both the inner copper diameter and the PU coating thickness were obtained from the manufacturer’s specifications (0.2030 mm and 0.0105 mm, respectively). Our results show that when the wires are subject to the peak voltage for VADs (~14.5 V) and a steady-state current of 1.2 A (Fig. 1b), the temperature increases 0.65°C in the core of the copper wires (Fig. 1c), due to the inrush current. Nonetheless, the surface temperature of the patch in the steady state remains around 0.02°C, indicating that there is no significant risk for skin injury from heat dissipation. This combination of experimental and computational findings will enable the design of new percutaneous medical devices to support therapies that require safe exchange of power, signal, and mass through the human body.de_CH
dc.language.isoende_CH
dc.publisherSSBEde_CH
dc.rightsLicence according to publishing contractde_CH
dc.subjectSkinde_CH
dc.subjectVentricular assistance device (VAD)de_CH
dc.subjectCOMSOLde_CH
dc.subjectBiophysikde_CH
dc.subjectHautde_CH
dc.subject.ddc610.28: Biomedizin, Biomedizinische Technikde_CH
dc.titleThermal analysis of new transdermal devices for power transfer to ventricular assist devicesde_CH
dc.typeKonferenz: Posterde_CH
dcterms.typeTextde_CH
zhaw.departementSchool of Engineeringde_CH
zhaw.organisationalunitInstitute of Computational Physics (ICP)de_CH
zhaw.publisher.placeBaselde_CH
dc.identifier.doi10.21256/zhaw-29026-
zhaw.conference.detailsAnnual Meeting of the Swiss Society for Biomedical Engineering (SSBE), Allschwil, Switzerland, 13 September 2023de_CH
zhaw.funding.euNode_CH
zhaw.originated.zhawYesde_CH
zhaw.publication.statuspublishedVersionde_CH
zhaw.publication.reviewPeer review (Abstract)de_CH
zhaw.webfeedSensors and Measuring Systemsde_CH
zhaw.author.additionalNode_CH
zhaw.display.portraitYesde_CH
Appears in collections:Publikationen School of Engineering

Files in This Item:
File Description SizeFormat 
2023_GrossmannAybar-etal_Thermal-analysis-of-new-transdermal-devices.pdf863.18 kBAdobe PDFThumbnail
View/Open
Show simple item record
Grossmann Aybar, C., Fehr, D., Spano, F., Kourouklis, A. P., Mazza, E., & Bonmarin, M. (2023, September 13). Thermal analysis of new transdermal devices for power transfer to ventricular assist devices. Annual Meeting of the Swiss Society for Biomedical Engineering (SSBE), Allschwil, Switzerland, 13 September 2023. https://doi.org/10.21256/zhaw-29026
Grossmann Aybar, C. et al. (2023) ‘Thermal analysis of new transdermal devices for power transfer to ventricular assist devices’, in Annual Meeting of the Swiss Society for Biomedical Engineering (SSBE), Allschwil, Switzerland, 13 September 2023. Basel: SSBE. Available at: https://doi.org/10.21256/zhaw-29026.
C. Grossmann Aybar, D. Fehr, F. Spano, A. P. Kourouklis, E. Mazza, and M. Bonmarin, “Thermal analysis of new transdermal devices for power transfer to ventricular assist devices,” in Annual Meeting of the Swiss Society for Biomedical Engineering (SSBE), Allschwil, Switzerland, 13 September 2023, Sep. 2023. doi: 10.21256/zhaw-29026.
GROSSMANN AYBAR, Christian, Daniel FEHR, Fabrizio SPANO, Andreas P. KOUROUKLIS, Edoardo MAZZA und Mathias BONMARIN, 2023. Thermal analysis of new transdermal devices for power transfer to ventricular assist devices. In: Annual Meeting of the Swiss Society for Biomedical Engineering (SSBE), Allschwil, Switzerland, 13 September 2023. Conference poster. Basel: SSBE. 13 September 2023
Grossmann Aybar, Christian, Daniel Fehr, Fabrizio Spano, Andreas P. Kourouklis, Edoardo Mazza, and Mathias Bonmarin. 2023. “Thermal Analysis of New Transdermal Devices for Power Transfer to Ventricular Assist Devices.” Conference poster. In Annual Meeting of the Swiss Society for Biomedical Engineering (SSBE), Allschwil, Switzerland, 13 September 2023. Basel: SSBE. https://doi.org/10.21256/zhaw-29026.
Grossmann Aybar, Christian, et al. “Thermal Analysis of New Transdermal Devices for Power Transfer to Ventricular Assist Devices.” Annual Meeting of the Swiss Society for Biomedical Engineering (SSBE), Allschwil, Switzerland, 13 September 2023, SSBE, 2023, https://doi.org/10.21256/zhaw-29026.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.