dc.description.abstractTestosterone, androstenedione and 17-hydroxyprogesterone are powerful steroids which play a critical role in development and reproductive processes in men and women. The quantitative determination of serum steroids is mandatory in the investigation of suspected disorder of excessive and insufficient steroid production in adults and delayed puberty in adolescents. While immunoassays are the general analytical method, for accurate and reproducible quantification of such analytes liquid chromatography coupled with tandem mass spectrometry (LC-MS/MS) is gaining importance. Advantages are its unparalleled specificity and sensitivity combined with the very practical potential for multi-analyte quantification. Prior to any analysis by LC-MS/MS a suitable sample clean-up is essential to efficiently remove matrix components such as proteins, lipids, carbohydrates and salt which may interfere with the analytical procedure. The Tecan AC Extraction PlateTM is an ingenious novel consumable: a 96-position, deep-well microplate featuring TICETM (Tecan Immobilized Coating Extraction) technology. The inner surface of each well is coated with a specific volume of sorptive material to approximately half the well’s height. The TICE coating acts as the extractant, eliminating the need for protein precipitation, liquid separation steps or the generation of emulsions. It is the basis for a simple and automatable approach to the preparation of biological samples without the need for offline operations. In parallel an automation for SLE+ plates from Biotage was set up on the same Tecan liquid handling system for comparison.de_CH
dc.language.isoende_CH
dc.publisherTecan Tradingde_CH
dc.rightsNot specifiedde_CH
dc.subject.ddc572: Biochemiede_CH
dc.titleSample preparation of multiple steroids for quantitative determination by LC-MS/MS : comparison of two extraction proceduresde_CH
dc.typeKonferenz: Posterde_CH
dcterms.typeTextde_CH
zhaw.departementLife Sciences und Facility Managementde_CH
zhaw.organisationalunitInstitut für Chemie und Biotechnologie (ICBT)de_CH
dc.identifier.doi10.21256/zhaw-4886-
zhaw.conference.detailsMSACL 2015 US Congress, San Diego, USA, March 28 - April 1 2015de_CH
zhaw.funding.euNode_CH
zhaw.originated.zhawYesde_CH
zhaw.publication.statuspublishedVersionde_CH
zhaw.publication.reviewNot specifiedde_CH
zhaw.funding.zhawInnovative Extraktionstechnologie zur automatisierten Probenvorbereitung für Massenspektrometrie-Applikationen im integrierten Laborde_CH
Appears in collections:Publikationen Life Sciences und Facility Management

Files in This Item:
File Description SizeFormat 
Poster_ZHAW-Tecan_MSACL2015.pdf1.34 MBAdobe PDFThumbnail
View/Open
Show simple item record
Sebregondi, P., Zaborosch, C., van Staveren, D., Geyer, R., & Svoboda, M. (2015). Sample preparation of multiple steroids for quantitative determination by LC-MS/MS : comparison of two extraction procedures. MSACL 2015 US Congress, San Diego, USA, March 28 - April 1 2015. https://doi.org/10.21256/zhaw-4886
Sebregondi, P. et al. (2015) ‘Sample preparation of multiple steroids for quantitative determination by LC-MS/MS : comparison of two extraction procedures’, in MSACL 2015 US Congress, San Diego, USA, March 28 - April 1 2015. Tecan Trading. Available at: https://doi.org/10.21256/zhaw-4886.
P. Sebregondi, C. Zaborosch, D. van Staveren, R. Geyer, and M. Svoboda, “Sample preparation of multiple steroids for quantitative determination by LC-MS/MS : comparison of two extraction procedures,” in MSACL 2015 US Congress, San Diego, USA, March 28 - April 1 2015, 2015. doi: 10.21256/zhaw-4886.
SEBREGONDI, Patrizia, Christiane ZABOROSCH, Dave VAN STAVEREN, Roland GEYER und Michal SVOBODA, 2015. Sample preparation of multiple steroids for quantitative determination by LC-MS/MS : comparison of two extraction procedures. In: MSACL 2015 US Congress, San Diego, USA, March 28 - April 1 2015. Conference poster. Tecan Trading. 2015
Sebregondi, Patrizia, Christiane Zaborosch, Dave van Staveren, Roland Geyer, and Michal Svoboda. 2015. “Sample Preparation of Multiple Steroids for Quantitative Determination by LC-MS/MS : Comparison of Two Extraction Procedures.” Conference poster. In MSACL 2015 US Congress, San Diego, USA, March 28 - April 1 2015. Tecan Trading. https://doi.org/10.21256/zhaw-4886.
Sebregondi, Patrizia, et al. “Sample Preparation of Multiple Steroids for Quantitative Determination by LC-MS/MS : Comparison of Two Extraction Procedures.” MSACL 2015 US Congress, San Diego, USA, March 28 - April 1 2015, Tecan Trading, 2015, https://doi.org/10.21256/zhaw-4886.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.