Publication type: Article in scientific journal
Type of review: Peer review (publication)
Title: Measurement and analysis of electromagnetic fields of pulsed magnetic field therapy systems for private use
Authors: Järmann, Thomas
Suter, Fabian
Osterwalder, Diego
Lüchinger, Roger
DOI: 10.1088/0952-4746/31/1/007
Published in: Journal of Radiological Protection
Volume(Issue): 31
Issue: 1
Page(s): 107
Pages to: 116
Issue Date: 2011
Publisher / Ed. Institution: IOP Publishing
ISSN: 0952-4746
1361-6498
Language: English
Subject (DDC): 530: Physics
615: Pharmacology and therapeutics
Abstract: Recently, pulsed magnetic field therapy (PMFT) systems have become available for private use. Although they may be applied without medical supervision, only a little is known about their field quantities. In this study, the spatial distribution and the temporal characteristics of the magnetic flux densities of three PMFT systems, available in Europe, were analysed. In close proximity to the surface, the maxima of the peak magnetic flux densities were 461 µT, 170 µT and 133 µT, respectively. At a distance of 30 cm above the whole body mat, the peak magnetic flux density was 77 µT. The excitation patterns consisted of repeating bursts with carrier frequencies between 210 and 1667 Hz. In conclusion, magnetic flux densities were far above International Commission on Non-Ionizing Radiation Protection reference levels. Since these systems are supposed to be medical devices as well as wellness devices, risk analysis of PMFT systems and the effectiveness of these devices need to be investigated in future studies.
URI: https://digitalcollection.zhaw.ch/handle/11475/4989
Fulltext version: Published version
License (according to publishing contract): Licence according to publishing contract
Departement: School of Engineering
Appears in collections:Publikationen School of Engineering

Files in This Item:
There are no files associated with this item.
Show full item record
Järmann, T., Suter, F., Osterwalder, D., & Lüchinger, R. (2011). Measurement and analysis of electromagnetic fields of pulsed magnetic field therapy systems for private use. Journal of Radiological Protection, 31(1), 107–116. https://doi.org/10.1088/0952-4746/31/1/007
Järmann, T. et al. (2011) ‘Measurement and analysis of electromagnetic fields of pulsed magnetic field therapy systems for private use’, Journal of Radiological Protection, 31(1), pp. 107–116. Available at: https://doi.org/10.1088/0952-4746/31/1/007.
T. Järmann, F. Suter, D. Osterwalder, and R. Lüchinger, “Measurement and analysis of electromagnetic fields of pulsed magnetic field therapy systems for private use,” Journal of Radiological Protection, vol. 31, no. 1, pp. 107–116, 2011, doi: 10.1088/0952-4746/31/1/007.
JÄRMANN, Thomas, Fabian SUTER, Diego OSTERWALDER und Roger LÜCHINGER, 2011. Measurement and analysis of electromagnetic fields of pulsed magnetic field therapy systems for private use. Journal of Radiological Protection. 2011. Bd. 31, Nr. 1, S. 107–116. DOI 10.1088/0952-4746/31/1/007
Järmann, Thomas, Fabian Suter, Diego Osterwalder, and Roger Lüchinger. 2011. “Measurement and Analysis of Electromagnetic Fields of Pulsed Magnetic Field Therapy Systems for Private Use.” Journal of Radiological Protection 31 (1): 107–16. https://doi.org/10.1088/0952-4746/31/1/007.
Järmann, Thomas, et al. “Measurement and Analysis of Electromagnetic Fields of Pulsed Magnetic Field Therapy Systems for Private Use.” Journal of Radiological Protection, vol. 31, no. 1, 2011, pp. 107–16, https://doi.org/10.1088/0952-4746/31/1/007.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.